Specifications |
CO2SMO® Plus!
Pulse Oximeter
Co2SMO
Plus!
Capnograph
- Principle of Operation
Solid state mainstream sensor. Single beam, non-dispersive infrared
(NDIR) absorption, ratiometric measurement.
- Initialization Time
Capnogram within 15 sec., full specifications within 60 sec.
- Calibration
Routine calibration is not required. 10 sec. adapter zero performed
when changing to different style of adapter.
- Accuracy Verification
Accuracy verifier provided on sensor cable.
- Capnostat® Mainstream CO2 Sensor
| Size: |
1.30" x 1.67" x .85" (3.3 x 4.2 x
2.2 cm) |
| Weight: |
0.64 oz. (18 grams) - cable excluded |
- Airway Adapters
Neonatal, pediatric and pediatric/adult single patients use Series
3 combined CO2 / flow and neonatal and pediatric/adult flow sensors.
Reusable and single patients use CO2 adapter also available.
- Sampling Adapter
Order Sampling Adapter Starter Kit (P/N 5816-34) for sidestream sampling
of non-intubated patients (pump draw 180 ml/min).
- CO2 (Carbon Dioxide)
| Compensations: |
Gas composition (selectable)
Barometric Pressure (automatic) |
| Capnogram: |
Selectable sweep speed |
| Range: |
0 - 150 mmHg, 0 - 19.7%, 0 - 20 kPa at 760 mmHg Pb |
| Accuracy: |
+/- 2 mmHg (for 0 - 40 mmHg)
+/- 5 % of reading (for 41 - 70 mmHg)
+/- 8 % of reading (for 71 - 1500 mmHg) |
| Resolution: |
1 mmHg |
| Response Time: |
Less than 60 ms |
| Accuracy: |
±1 br/min |
| Range: |
0-150 br/min |
| Resolution: |
1 br/min |
Pulse Oximeter
- Principle of Operation
Red/Infrared absorbtion
| Range: |
0 - 100% |
| Accuracy: |
+/- 2% SpO2 (for 80 - 100% SpO2)
(1 SD, or 68% of readings within claim)
Unspecified (for 0 - 79% SpO2) |
| Resolution: |
1% |
| Averaging: |
Menu selectable, 2 or 8 seconds |
| Audio: |
Pitch of pulse tome varies with SpO2 value |
| Range: |
30 - 250bpm |
| Accuracy: |
+/- 1 % of full scale |
| Resolution: |
1 bpm |
| Averaging: |
8 seconds |
- Sensors
Reusable finger and Y-Sensor™
CO2MO Plus!
- Parameter
Mechanical, spontaneous and total MV and MV alveolar, Vt(i & e),
Mechanical Vt/ml?kg, Vt alveolar and Vd airway. VCO2,
VCO2/kg, PECO2 (FeCO2) and ETCO2.
Mechanical, spontaneous and total RR, SpO2 and PR. Raw
(i & e), Cdyn, Cstatic, PIP, NIP, MAP, PEEP, AUTO PEEP, PIF, PEF,
Pplat, I:E, ti, te and RSBI. Physiological deadspace (Vdphy), alveolar
deadspace (Vdalv) and physiologic Vd/Vt calculated with user entry
of PaCO2
- Trend Memory
28 parameters trended for 24 hours, all parameter available
through serial outputs. VCO2 and mechanical, spontaneous
and total MV alv trended on-screen.
| Output: |
To a PC for printing, enhanced graphic and trending,
and also to patient monitoring and information systems |
| Input: |
Receives data from various ABG monitors for automatic
Vd/Vt phys, Vd alv and Vd phys calculations |
- Alert
Adjustable limits for ETCO2, SpO2, RR, PR, & No Respiration
| Audio: |
Adjustable volume, 2 min. silence or OFF (LED indicators) |
| Visual: |
On-screen indicator and red "Alert Bar" |
| Type: |
Dot matrix, Cold Cathode Display (CCD |
| Size: |
1.5" H x 5" W (3.81 x 12.7 cm) |
| Size: |
3.3" H, 9.0" W, 8.0" D (8.38 x 22.86
x 20.32 cm) |
| Weight: |
8 lb (3.63 kg) |
| Power requirements: |
100 - 120/200 - 240 VAC, 50 - 60 Hz, 30 VA |
| Battery: |
8 lb (3.Sealed lead acid gel cell, 2 hr. life,
12 hr. recharge kg) |
| Operating temperature: |
50 - 104°F (10 - 40°C) |
| Operating humidity: |
0 - 90% relative humidity (non-condensing) |
|